LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Rear Case Replacement
319
8
10. Install four battery pins as described in
Battery Pin Replacement (p. 364)
). Place the
bottom of energy storage capacitor flush to the rear case.
12. Route the energy storage capacitor wires along left side of capacitor body.
NOTE:
Place transfer relay on top of the energy storage capacitor until further in the reassembly process.
13. Install the biphasic PCB as described in
Installing the Biphasic Module (A22)/Inductive Resistor (A14) (p. 302)
14. Install the parameter bezel as described in
Installing the Parameter Bezel (p. 313)
.
15. Install the power PCB as described in
Installing the Power PCB (A03) (p. 257)
16. If option is present, install the SpO2 module as described in
Installing the SpO2 PCB (A16) (p. 280)
.
17. Install the NIBP/CO2 Module housing as described in
Installing the NIBP (A21)/CO2 (A23) Modules (p. 292)
18. Install the transfer relay assembly as described in
Installing the Transfer Relay Assembly (A13) (p. 270)
19. Install the OEM PCB as described in
Installing the OEM PCB (A06) (p. 265)
20. Install the system/therapy PCB assembly as described in
System (A01)/Therapy (A04) PCB Assembly Replacement (p. 232)
.
21. Reassemble the case as described in
Reassembling the Case (p. 184)
.
To install the rear case:
(Continued) 23 steps, (Page 2 of 3)