LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Power/Contact PCB Cable (W05) Replacement
327
8
Installing the Power/Contact PCB Cable (W05)
To install the power/contact PCB cable:
1.
Connect the power/contact PCB cable (
) to J42 of contact PCB (A07).
2.
Install the NIBP/CO2 assembly as described in
Installing the NIBP (A21)/CO2 (A23) Modules (p. 292)
.
3.
Install the OEM PCB as described in
Installing the OEM PCB (A06) (p. 265)
4.
Install the transfer relay assembly as described in
Installing the Transfer Relay Assembly (A13) (p. 270)
5.
Connect the power/contact PCB cable to the power PCB (A03) at J12 (may be labeled P12).
6.
Install the system/therapy PCB assembly as described in
Installing the System (A01)/Therapy (A04) PCB Assembly (p. 239)
7.
Reassemble the case as described in
Reassembling the Case (p. 184)
.