LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Installing the Speed Dial Assembly (W15)
220
8
Installing the Speed Dial Assembly (W15)
To install the speed dial assembly into the front case (
):
NOTE:
When installing a new speed dial assembly (
) use
External Hardware Repair Kit (MIN 3305431-016) (p. 507)
.
1.
Insert the speed dial assembly from inside the front case so that the
cable is extending away from the perimeter seal.
2.
Attach the speed dial assembly shaft to the front case with the nut and
lock washer (torque to 10 in-lb using a 9/16 deep socket).
3.
Push the knob (
) onto the speed dial assembly shaft
from the outside of the front case (see
4.
If interface PCB is present, connect the speed dial assembly to the
interface PCB at J33.
5.
Reassemble the case as described in
Reassembling the Case (p. 184)
.
rear of switch, orient with
cable toward PCB
Figure 8.28—Speed dial assembly
cable connection
to J33
Figure 8.29—Speed dial cable