LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
169
8
Replacement Procedures
The replacement procedures are a set of detailed instructions for disassembly, handling, and reassembly of replaceable LIFEPAK 15 monitor/
defibrillator parts. Perform an interior inspection whenever the LIFEPAK 15 monitor/defibrillator case is opened for service. When disconnecting
cables and wire harnesses, label the cables and connections so that they match easily during reassembly, for example, J01, J03, and so forth.
See the
for additional information.
Before replacing any parts, review the following items:
•
Summary of Replacement Procedures (p. 170)
•
Warnings and Cautions (p. 173)
•
Static-Sensitive Device Handling (p. 174)
•
•
Capacitor Discharge Tool (p. 177)
•
Capacitor Discharging Procedure (p. 178)
•
Discharging the C15 Pacing Capacitor (p. 179)
•
Saving and Restoring the Setup Configuration (p. 180)
•
Disassembling the Case (p. 181)
•
Reassembling the Case (p. 184)
•
Inside Front Case Diagram (p. 189)
•