LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Disassembling the Case
181
8
Disassembling the Case
Refer to
Front Parts Diagrams and Parts List (p. 377)
Rear Diagrams and Parts List (p. 399)
To disassemble the case halves:
1.
If applicable, remove carrying case.
2.
Remove all cables and patient connections, and then remove
the batteries.
3.
Remove the four front mounting screws (
the bumpers.
4.
Remove the four rear mounting screws (
the bumpers.
5.
Remove the bumpers (
)
from the case and set aside.
Figure 8.4—Bumper screw locations
Front bumper screws