LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Disassembling the Case
183
8
8.
Disconnect the P2 system/interface flex cable connector
(W04) (
) from the system PCB (see
9.
Disconnect the P23 therapy ribbon cable connector (W11)
(
) in the front case from the therapy PCB (see
).
10. Disconnect the J24 therapy cable connector (W11) from the
P24 transfer relay connector (A13) (see
).
11. Separate the front and rear cases halves.
12. Before continuing any further, discharge the energy storage
capacitor using the discharge tool (see
Figure 8.6—Disconnection points
To disassemble the case halves:
(Continued) 12 steps, (Page 3 of 3)
Disconnect P2
flex cable
Disconnect P23
ribbon cable
Disconnect J24
therapy cable
Lift connector
tab to
disconnect
cable