LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Counters
117
5
Understanding the Counters
The Service/Status/Counters overlay displays the counters shown in
Resetting the Counters
Select CLEAR ALL on the
Service/Status/Counters overlay to reset the subtotal counters in the boxes, but not the running-total
counters.
Table 5.4—Counters
Data
Description
Total Shocks
Running total of all the shocks ever delivered by the device. This counter
cannot be reset.
360 J Shocks
The number in the box represents the number of 360-joule shocks delivered
since the last reset. The number in the right column is a running total of all
360-joules shocks ever delivered by the device (cannot be reset).
225 - 325 J Shocks
The number in the box represents the number of 225-joule to 325-joule
shocks delivered since the last reset. The number in the right column is a
running total of all 225-joule to 325-joule shocks ever delivered by the
device (cannot be reset).
1 - 200 J Shocks
The number in the box represents the number of 0-joule to 200-joule shocks
delivered since the last reset. The number in the right column is a running
total of all 0-joule to 200-joule shocks ever delivered by the device (cannot
be reset).