LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
SpO2 Connector Cable (W22) Replacement
345
8
Removing the SpO2 Connector Cable (W22)
To remove the SpO2 connector cable (
) from the rear case (refer to
Inside Rear Case Diagrams (p. 228)
)
:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
System (A01)/Therapy (A04) PCB Assembly Replacement (p. 232)
.
3.
Remove the parameter bezel as described in
Removing the Parameter Bezel (p. 311)
.
4.
Remove the four screws (
) (using a Torx T-10 bit) securing the ferrite housing (
) and SpO2
connector cable. Discard the screws.
5.
Remove the connector from the parameter bezel.
Installing the SpO2 Connector Cable (W22)
To install the SpO2 connector cable (
) into the rear case (refer to
Inside Rear Case Diagrams (p. 228)
)
:
NOTE:
When replacing the SpO2 cable use
MASIMO Sp02 Panel Mount Cable Repair Kit (MIN 3305431-026) (p. 514)
1.
Fold SpO2 connector cable (W22) over the top of the ferrite bead and then position the ferrite bead behind the connector.
2.
) over the flex cable and ferrite bead.
3.
Insert the SpO2 connector cable into the parameter bezel and secure with the four new screws (
); torque to 2 in-lb
using a Torx T-10 bit (see
4.
Install the parameter bezel as described in
Installing the Parameter Bezel (p. 313)
.
5.
Install the system/therapy PCB assembly as described in
Installing the System (A01)/Therapy (A04) PCB Assembly (p. 239)
6.
Reassemble the case as described in
Reassembling the Case (p. 184)
.