LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Troubleshooting Chart
105
5
CO2 Feature
Note:
The CO2 module can
take up to 6 minutes for all
internal processes to
complete.
CO2 fails calibration
Verify test setup and retry test.
Check to see if CO2 calibration gas canister is empty.
Check FilterLine to see if it is disconnected.
Check for pinched hose inside device.
Possible A23 CO2 module failure.
CO2 displays FilterLine Blockage
message
Replace FilterLine.
Possible occlusion or kinking of input tubing between CO2 connector and CO2
module.
Possible A23 CO2 module failure.
CO2 displays XXX in the CO2 region of
the display with Service LED OFF
Replace FilterLine.
Possible occlusion or kinking of tubing inside device.
Possible A23 CO2 module failure.
CO2 displays XXX in the CO2 region
with Service LED ON (review diagnostic
log).
Review error code 9205
in
CO2 FilterLine installed with no
indication on display that CO2 is
connected
Clean the CO2 input connector with compressed air to remove loose debris.
Replace FilterLine.
Table 5.1— Corrective Actions (Continued)
Area
Observed Symptom
Suggested Corrective Action