LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Patient Parameter Service Codes
129
5
Patient Parameter Service Codes
Table 5.12—Initial Digit 9, Patient Parameter Service Codes (PP)
Code
Service Code Description
Corrective
Action Code
900e
PPSP_ERROR_CONFIG (SpO2 module detected but not in manufacturing configuration; param: 1 = found Masimo but should not, 2 =
unit not found.)
902d
PPSP_ERROR_MNC_CONFIG_B (SpO2 module detected but manufacturing configuration, not configured for Nellcor sensor support;
param: 1 = Masimo doesn’t support config B, 2 = Device doesn’t support config B.)
9119
PPNI_ERROR_CONFIG_MISMATCH (Device with NIBP module not configured for it; param = SW version.)
9205
PPCO_ERROR_WRITING_TO_MODULE (CO2 module detected but not in manufacturing configuration. Unit disabled. Param = config-
info.)
9304
PPTI_CONFLICT_IP_TEMPER_ERROR_CONFIG (IP and Temperature manufacturing configuration bits both on at the same time.)