LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Printer Assembly/Chassis Ground Cable (W19) Replacement
227
8
Printer Assembly/Chassis Ground Cable (W19) Replacement
Refer to
Inside Front Case Diagram (p. 189)
Removing the Printer Assembly/Chassis Ground Cable (W19)
To remove the printer/chassis ground W19 cable
) (see
from the front case:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the screw (
) securing the W19 cable to the interface PCB and LCD
display assembly bracket (
). Discard the screw.
3.
Disconnect the W19 cable ground terminal from the printer assembly connector bracket.
Installing the Printer Assembly/Chassis Ground Cable (W19)
To install the printer/chassis ground W19 cable
) into the front case:
1.
Connect the W19 cable ground terminal to the printer assembly connector bracket.
2.
Secure the W19 cable to the interface PCB and LCD display assembly bracket (
) with one new screw (
); torque to 6.8 in-lb.
3.
Reassemble the case as described in
Reassembling the Case (p. 184)
.
chassis ground
Printer Assembly (A12)
Figure 8.34—printer cable
connection