LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Biphasic Module (A22)/Inductive Resistor (A14) Replacement
302
8
Installing the Biphasic Module (A22)/Inductive Resistor (A14)
To install the biphasic PCB and inductive resistor:
NOTE:
When replacing the Biphasic module, use
Repair Kit (MIN 3305431-011) (p. 504)
.
1.
) to the biphasic bracket
). Tilt the PCB to get into the bracket slots. (Do not
bend the PCB).
2.
Wrap the inductive resistor (
four turns and connect to the biphasic PCB at J102 and J108.
Ensure that the twisted wires are routed below and around CR7.
3.
(If removed) Clip the inductive resistor (
) into the
biphasic bracket with the wires exiting away from the bracket.
4.
PCB at J105.
5.
PCB at J103.
6.
PCB at J101.
7.
If inductive resistor was removed: Install two new screws with washer
(
) into the bracket securing the inductive resistor in
place; torque to 4.0 in-lb.
Figure 8.96—Biphasic PCB installation
J102
J105
inductive resistor screws
and washers (2 places)
J108
J103
J101
J104