LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
SpO2 PCB (A16) Replacement
278
8
11. To remove the SpO2 flex cable, remove the one
screw (
) attaching the SpO2/OEM flex
) from the SpO2
module. Discard the screw.
To remove the SpO2 PCB from the rear case (see
Figure 9.21 on p. 413
):
(Continued) 13 steps, (Page 4 of 5)
flex cable
screw
location
Figure 8.70—SpO2 flex cable screw