LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Reassembling the Case
188
8
NOTE:
If there is no indication of power, either the batteries are dead or no batteries are installed, or the W04 cable connection
was not made when the case was reassembled. The W04 cable is the flex cable between the front case and rear case.
Check the batteries and, if they are charged, disassemble the case as described in
Disassembling the Case (p. 181)
and
check the W04 cable.
9.
Install the batteries.
NOTE:
Pay special attention to the
SERVICE
indicator as you turn on the device in the next step.
10. If applicable, reinstall carrying case.
11. Turn the device ON and observe the SERVICE indicator.
•
If the SERVICE indicator is OFF, continue with step 12.
•
If the SERVICE indicator is ON, skip to step 13.
•
If the device gives no indication that power is on, review information in Note below.
12. Complete the Test and Calibration Procedures, followed by the Performance Inspection Procedures.
13. Error codes have been written to the Service Log. Continue to
Processing Service Log Codes (p. 115)
To reassemble the case halves:
(Continued)
13 steps,
(Page 5 of 5)