Installing the System (A01)/Therapy (A04) PCB Assembly
252
8
8.
Follow the instructions to install the system/therapy assembly into device as described in
Installing the System (A01)/Therapy
.
9.
After device reassembly, complete the TCP – Defibrillator Energy Calibration procedure. The defibrillator calibration constants
may be invalidated when you replace Therapy PCB.
To replace the therapy PCB (A04):
(Continued) 9 steps, (Page 3 of 3)