LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Tools List
176
8
Tools List
The suggested list of tools for the LIFEPAK 15 monitor/defibrillator replacement procedures is as follows:
•
Static-dissipative mat and wrist strap
•
Anti-static rack and/or conductive bags
•
Capacitor discharge tool (information about
Capacitor Discharge Tool (p. 177)
•
Torque screwdriver(s)—required torque settings are 2.0, 4.0, 6.8, and 10.0 in-lb
•
Point 0 power drive bit (P0)- Phillips tip for 2-56 screws
•
Point 1 power drive bit (P1)- Phillips tip for 4-40 screws (shaft length of 2” and 6”)
•
Point 2 power drive bit (P2)- Phillips tip for 6-32 screws (shaft length of 6”)
•
T10 torx power drive bit
•
T15 torx power drive bit
•
4 mm modified thin wall socket for battery pins (Physio-Control tool #3300955)
•
1/4” socket
•
1/2” deep socket
•
9/16” deep socket
•
3/4” modified deep socket (Physio-Control tool #3305448)
•
3/8” nutdriver
•
Small, slotted screwdriver for label removal
•
Diagonal cutter
•
Needle nose pliers
•
1.5” (hard) putty knife
•
See PIP–Test Equipment Requirements and TCP–Test Equipment Requirements for additional test tools required for
performance testing
Torque in-lb
2.0
4.0
6.8
10.0
Torque in-oz
32
64
108.8
160
Torque cNm
22.5
45
77
113