LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Installing the System (A01)/Therapy (A04) PCB Assembly
249
8
6.
Gently lift the therapy PCB up
and away from the system PCB.
The two PCBs are linked by the
30-pin header, which is a direct-
connection contact assembly.
7.
Remove the 30-pin header
assembly from the therapy board
(
).
8.
Remove the three retaining
screws (
) on the
therapy board. Discard the
screws.
9.
Remove the EMI shield.
To remove the therapy PCB (A04):
(Continued) 9 steps, (Page 3 of 3)
Figure 8.49—30-pin header (therapy PCB side)
Figure 8.49—Therapy EMI shield and screw locations
screws
(3 places)
EMI shield