LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Reassembling the Case
185
8
4.
Carefully connect the P2 system/interface flex cable
connector (W04) (
5.
Ensure the case seal is in place in the groove along outside
edge of front case. Fold the front and rear case halves
together. Route the therapy cable wires away from the edge of
the case to avoid pinching.
CAUTION
POSSIBLE COMPONENT DAMAGE
- Ensure the therapy cable
wires are routed away from the edge of case to avoid wire pinching
(see
).
POSSIBLE MOISTURE LEAKAGE
- Visually inspect the mating
surfaces between the front and the rear case halves before and after
fastening them together to ensure that they are even.
PREVENT VIBRATION DAMAGE
In order to meet vibration
specifications (for example, prevent loosening of case screws), use new
screws (
) when assembling the case and torque to
specified value.
To reassemble the case halves:
(Continued)
13 steps,
(Page 2 of 5)
Connect P2
flex cable
Connector tab
During closure, route
therapy wires away
from case edges
Figure 8.7—Connection points