LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Therapy Service Codes
135
5
a204
PA_ERROR_2MS_OVERRUN (2 ms ECG sampling overrun.)
a208
PA_ERROR_MSG_RESYNC (Received message incomplete.)
a209
PA_ERROR_MSG_SIZE (Received msg size error/input buff full.)
a20d
PA_ERROR_PACE_OVERRUN (Pacing pulse process overrun. Set current to zero.)
a20e
PA_ERROR_PULSE_WIDTH (Pacing pulse too long.)
a20f
PA_ERROR_A2D_INT (Internal A/D conversion timeout. Set current to zero.)
a210
PA_ERROR_A2D_EX (External A/D conversion timeout. Current set to zero.)
a211
PA_ERROR_SPI (SPI transfer timeout. Current set to zero.)
a212
PA_ERROR_RAM_TEST (RAM test failed. Reset Pacer Processor.)
a213
PA_ERROR_ROM_TEST (ROM CRC test failed.)
a214
PA_ERROR_CPU_TEST (Stack overrun occurred.)
a215
PA_ERROR_STACK_CHECK (Is5 V ref. out of range.)
a216
PA_ERROR_V_ISO_MON (HV present when not pacing.)
a217
PA_ERROR_V_12V_MON (+12 V voltage out of range.)
a218
PA_ERROR_V_HVIS_SENSE (HV present when not pacing.)
Table 5.13—Initial Character a, Therapy Service Codes (TH)
Code
Service Code Description
Corrective
Action Code