LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Invasive Pressure Connector Assembly (W33) Replacement
356
8
Removing the Invasive Pressure Connector Assembly (W33)
To remove the invasive pressure connector assembly from the rear case (refer to
Inside Rear Case Diagrams (p. 228)
)
:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Disconnect the invasive pressure cable (
) from the J7 connector on the system PCB (A01).
3.
Remove the parameter bezel as described in
Removing the Parameter Bezel (p. 311)
.
4.
Remove the eight screws and captured washers (
) that secure the IP connectors to the parameter bezel. Remove
the IP connectors and gaskets.
5.
Remove the IP connectors and gaskets. Discard the screws and gaskets.
Installing the Invasive Pressure Connector Assembly (W33)
To install the Invasive Pressure Connector Assembly into the rear case (refer to
Inside Rear Case Diagrams (p. 228)
)
:
NOTE:
When installing a new invasive pressure connector, use the
Invasive Pressure Connector Repair Kit (MIN
.
1.
Secure the IP connectors and new gaskets (
)
to the parameter bezel with the eight new screws (
torque to 6.8 in-lb. (See
for connector orientation.)
2.
Install the parameter bezel as described in
Installing the Parameter Bezel (p. 313)
.
3.
Connect the IP connector cable (
) to the system PCB (A01) at J7.
4.
Reassemble the case as described in
Reassembling the Case (p. 184)
.
CAUTION
POSSIBLE MOISTURE LEAKAGE
When installing the cable, use new gaskets to help prevent ingress of fluids.