LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Power/Contact PCB Cable (W05) Replacement
326
8
Power/Contact PCB Cable (W05) Replacement
Power/contact PCB cable replacement consists of the following procedures:
•
Removing the Power/Contact PCB Cable (W05) (p. 326)
•
Installing the Power/Contact PCB Cable (W05) (p. 327)
Removing the Power/Contact PCB Cable (W05)
To remove the power/contact PCB cable (
) from the rear
case:
1.
Disassemble the case as described in
.
2.
Remove the system/therapy PCB assembly as described in
(A01)/Therapy (A04) PCB Assembly Replacement (p. 232)
.
3.
Disconnect the power/contact PCB cable from the power PCB at
J12 (may be labeled P12).
4.
Remove the transfer relay as described in
.
NOTE:
It is not necessary to remove the spade terminals as
described in the procedure.
5.
Remove the OEM PCB as described in
6.
Remove the NIBP/CO2 assembly as described in
NIBP (A21)/CO2 (A23) Modules (p. 288)
7.
Disconnect the power/contact PCB cable at J42 of the contact PCB (A07).
8.
Remove the power/contact cable.
P42
P12
Power PCB (A03)
A07
Contact PCB
Figure 8.107—Power/contact PCB
cable connections