LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
OEM PCB (A06) Replacement
262
8
Removing the OEM PCB (A06)
Refer to
Inside Rear Case Diagrams (p. 228)
To remove the OEM PCB (A06) from the rear case
:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as
described in
System (A01)/Therapy (A04) PCB
.
3.
Disconnect Power/Therapy Cable (W02) (
) from J8 on the power PCB.
4.
Remove the OEM shield (
W02 - power/
therapy cable
OEM shield
W01 - power/
system cable
Figure 8.59—OEM PCB shield