LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Therapy Service Codes
132
5
a02d
TH_ERROR_WRONG_DEFIB_TYPE (Wrong Defib. type for software.)
a02e
TH_ERROR_ADC_READ (ADC read failure when getting cap charge.)
a101
DE_ERROR_LONG_WDT (Long watchdog test failed or watchdog did not reset in time.)
a102
DE_ERROR_SHORT_WDT (Short watchdog test failed.)
a103
DE_ERROR_SCI_RCV (Serial port receiver error.)
a104
DE_ERROR_XFER_KEY (Defib HW error.)
a106
DE_ERROR_ENERGY_OUT_OF_BOUND (VCAP-1 over/under charge.)
a107
DE_ERROR_HP_ENG_SELECT (Cannot determine the rotary setting.)
a109
DE_ERROR_CAL_CRC (Calibration Table CRC error.)
a10b
DE_ERROR_CHG_TIME (Takes too long to reach charge.)
a10c
DE_ERROR_CHG_ENABLE (CHG_EN1 is stuck high.)
a10d
DE_ERROR_DUMP_ENERGY (Defib HW error.)
a10e
DE_ERROR_RCV (Defib HW error.)
a111
DE_ERROR_ENERGY_NOT_ZERO (Unexpected energy while biphasic is in high-pot test.)
a112
DE_ERROR_TEST_XFER_ENABLE (Problem with transfer; turn on at main.)
Table 5.13—Initial Character a, Therapy Service Codes (TH)
Code
Service Code Description
Corrective
Action Code