LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Power PCB (A03) Replacement
260
8
10. Route CO2 adapter cable (W30) above the power/
system cable as shown.
11. Connect the power/system cable (W01) at J17.
12. Install the OEM PCB as described in
13. Install the system/therapy PCB assembly as
System (A01)/Therapy (A04) PCB
.
14. Reassemble the case as described in
.
To install the power PCB (A03) into the rear case:
(Continued) 14 steps, (Page 4 of 4)
J17
connection
Figure 8.58—Power/system PCB cable (W01)
CO2 adapter cable