LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Power PCB (A03) Replacement
255
8
6.
assembly (
). Discard the screws.
7.
Lift the power PCB and bracket partially out of the rear case.
Use caution not to strain the system connector cable.
8.
Disconnect the two W08 - system cable connectors at J9 and
J10 from the power PCB.
9.
If you are replacing the power PCB, remove the six screws
(
) securing the PCB to the mounting bracket
), and then remove the PCB. Discard the
screws.
To remove the power PCB (A03) from the rear case:
(Continued) 10 steps, (Page 2 of 3)
mounting screws
(2 places)
(V2 pictured) screws mounting
PCB to bracket (6 places)
J10
connection
J9
connection
Figure 8.52—Power PCB
screw locations