NEMA Performance Measurement
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
10-11
Procedure Details
When both the X and Y measurements have
been completed, a pop-up similar to the one
shown on the right appears.
14.
Note the results and click OK to close the
pop-up.
Resolution specifications are:
UFOV FWHM
≤
3.7
mm
CFOV FWHM
≤
3.7
mm
UFOV FWTM
≤
7.6
mm
CFOV FWTM
≤
7.6
mm
Linearity specifications are:
UFOV absolute
≤
0.5
mm
CFOV absolute
≤
0.5
mm
UFOV differential
≤
0.2
mm
CFOV differential
≤
0.2
mm.
15.
Replace the NEMA Y line phantom with
the collimator.
16.
Click Exit to return to the NEMA menu.
Note
. To measure the other detector, place the CardioMD system in the patient load position be-
fore repeating the procedure outlined above.
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......