CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-63
Procedure Details
At this point, it is necessary to adjust the entire
assembly in order for the shutter indicator /
manual lever to be able to move smoothly.
24.
Check that the edge of the actuator bracket
is perpendicular to the slide rail in the
bottom of the line source compartment.
25.
Check that the shutter indicator / manual
lever moves unhindered through the
opening in the line source compartment.
26.
Tighten the four Allen screws holding the
shutter indicator / manual lever.