Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-11
8.1.4
Ceiling Height
To provide sufficient service access, permit NEMA standard flood testing, and allow for proper
air circulation, the minimum allowable ceiling height should be 8’ 0” (245 cm) A/FF. 9’ 0”
(275 cm) A/FF ceiling height is recommended. (A/FF signifies ‘above finished floor’.)
8.1.5
Ancillary Services
The ancillary services for the system should also be considered, when determining the floor space
allocation. This would include areas such as:
•
Reception/waiting room
•
Darkroom
•
Hot lab
•
Storage
•
Dressing rooms
•
Toilets
•
Physician’s office, etc.
8.1.6
Miscellaneous
8.1.6.1
General
While planning the site, keep in mind that storage space for supplies, such as linens, on-site spare
parts, film, medical supplies, etc., is needed.
8.1.6.2
Floor
The room floor must consist of a finished concrete floor covered with an industrial-grade anti-
static tile. No carpet may be used on the floor.
8.1.6.3
Lighting
Lighting fixtures, vents, etc. shall be flush with finished ceiling.
Dimmers should be used to control light levels in control and exam rooms. Locate dimmers
conveniently for gantry and controls.
8.1.6.4
Housekeeping
Housekeeping rules, which should be observed in the nuclear room, include:
•
Never use steel wool for cleaning in any of the rooms. The fine metal fibers can enter
enclosures via air cooling systems and cause internal short circuits that are difficult to detect
and correct, causing equipment downtime.
•
Buffing of wax on floors can circulate a fine powder which could collect inside the
components on open contacts. Consequently, only liquid waxes should be used.
Note
. Past experience indicates that it is advantageous to install a sink in the nuclear room.
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......