Diagnostics
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
5-23
Procedure Details
4.
Measure the gap in both collimator locks
as shown in the photo on the right.
This gap should be 1.5 – 2.0 mm.
5.
Check that all five ball contacts are
engaging at the same point.
The contact balls must be evenly soldered and
none of the boards must be bent.
6.
Push in the collimator all the way. Check
that both collimator locks are fully
engaging (no gap).