CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-31
Procedure Details
The contents of the
ACoption.ini
file
determine whether the AC option is active or
not. The arrow in the figure on the right
indicates the entry to look for.
15.
To disable the CardioMD AC Option, edit
the entry to say
ACoption=0
.
This setting allows the CardioMD system to
function without the AC option.
16.
Select
File
,
Save
to save the modified file.
17.
Click the Close button to close down the
Notepad program.
Powering Down the Gantry
Procedure Details
Close down the acquisition PC and power off
the CardioMD gantry:
18.
Follow the system shutdown procedure
provided in Chapter 3
Getting Started
of
the CardioMD Operator’s Manual.
Detaching the Transmission Scanner
Procedure Details
19.
Unscrew and disconnect the cable to the
transmission scanner from the socket on
the detector.
CardioMD configuration file – AC option disabled
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......