Acquisition PC Software Updates
7-22
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
communication is working properly. In-
structions are provided in Chapter 5
Diag-
nostics
.
Installing the CardioMD Application Software
Procedure Details
1.
Insert the CardioMD software CD in the
acquisition PC’s CD-ROM drive.
2.
Close all programs on the acquisition PC.
3.
Start the Windows Explorer and select the
CD-ROM drive.
4.
Run the file
InstallCardioMDto-
tal.bat
by double-clicking on the name
of the file.
The file is located in the root directory of the
CD-ROM.
5.
Follow the instructions given.
6.
When the software installation is com-
pleted, restore calibration tables and cus-
tomer specific protocols from the backup
disks. Instructions are provided in the Ap-
pendix of the CardioMD Operator’s Man-
ual.
7.
Finally, remove the software CD-ROM
and store it in a safe place along with the
backup disks.
8.
Start the CardioMD acquisition software
by double-clicking the CardioMD icon on
the acquisition PC’s desktop.
7.3.3
Configuring Antivirus Software
Acquisition PC’s running Windows XP (CardioMD Series III systems) are supplied with antivirus
software. This section provides instructions for configuring Symantec AntiVirus™ on the acquisi-
tion PC.
If the CardioMD acquisition PC has no access to the internet or outside networks, you may want
to prevent it from trying to automatically check for new antivirus updates through its LiveUpdate
feature. To disable LiveUpdate, follow the instructions below.
III
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......