Repair Procedures
12-38
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
12.5.3
Line Source Compartment Repairs
12.5.3.1
Removing Line Source Compartment Covers
To access the parts inside the line source compartment, both covers must be removed. The
following sections explain how to remove covers:
•
While the transmission scanner is mounted on the detector, and
•
When the transmission scanner is dismounted and the line source compartment has been
detached from the AC option bracket.
Transmission Scanner Mounted on Detector
Procedure Details
1.
Use the hand controller to position the
detector assembly conveniently for access
to the transmission scanner you are going
to work with.
2.
If required, use the crank supplied with the
CardioMD system to position the line
source compartment conveniently.
Caution
. Make sure that all objects are clear of the detectors and transmission
scanners before activating hand controlled motions.
CAUTION
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......