Installation
12-92
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
5.
Attach the AC Option’s power interface
board (9ACP1316) to the detector power
supply.
6.
Plug the detector power cable into the
mating connector on the power interface
board.
7.
Attach the AC Option’s detector interface
cable (9CBL1313) to the DB15 connector
on the power interface board.
8.
Attach the AC Option’s CANbus interface
cable (9CBL1314) to the 6-way connector
on the power interface board.
CANbus cable
Detector interface cable