Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-63
Procedure Details
6.
Use a part of the rubber protector supplied
with the system to protect the cable from
the wall to the gantry.
3 m (10 ft) of rubber cable protectors are
provided to cover cables running from the
gantry to the wall (power and network cables)
and the cable between gantry and PC. The
cable protectors should be applied where
traffic across cables must be expected.
7.
Cut cable protectors to desired lengths.
Open the membrane at the base of the
protector to slide in the cable.
Connect the gantry-PC interface cable to the gantry as follows:
Procedure Details
8.
Push the cable coming from the
acquisition PC through the entry hole in
the gantry.
9.
Fasten the plate to the gantry using the
M4 x 8 screws.
10.
Connect the hand-controller/LED/E-stop
cable to the DB25 connector on the
internal connector panel.
11.
Connect the power cable to the IEC
connector.
12.
Connect the FireWire cable to any of the
ports on the FireWire hub.
DB25 connector
IEC connector
FireWire hub
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......