Acquisition PC Software Updates
7-20
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
dioMD
.
If the menu item says
Log off Administrator
:
6.
Click Cancel and proceed to the next sec-
tion.
Otherwise:
7.
Click OK.
The Log on to Windows dialog appears.
8.
In the User name field, enter
Adminis-
trator
. Ensure that the Password field is
left blank.
9.
Click OK.
Backing Up the Existing Software Version
Note
. As an extra precaution, if the CardioMD software is already installed on the acquisition PC,
the existing version can be saved by copying it to a backup directory. This ensures that customized
protocols, etc. are saved.
To back up the existing version of the CardioMD application software, copy the
C:\Cardiocam
directory to a working directory on the C: drive. Proceed as follows:
Procedure Details
1.
In the Windows Explorer’s folder list,
locate the
Cardiocam
directory.
2.
Right-click on the directory name and
select Copy from the pop-up menu.
3.
In the folder list, select the root directory,
C:\.
4.
Right-click in the file list, and select Paste
from the pop-up menu.
The system will create a new directory named
Copy of Cardiocam
.
Installing the FireWire driver on Windows 2000
Note
. This section applies to acquisition PC’s running Windows 2000.
If the CardioMD acquisition software has never been installed on the PC, the FireWire driver must
be installed.
Procedure Details
1.
Insert the CardioMD software CD in the
acquisition PC’s CD-ROM drive.
2.
Close all programs on the acquisition PC.
I, II
III
I, II
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......