CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-79
Procedure Details
6.
Again, press down each collision sensor at
the other end of the line source
compartment, one by one, easing the cover
into place.
7.
Remount the edge cover and fasten it
using the two Torx screws previously
removed from the service cover.
Transmission Scanner Dismounted
Procedure Details
1.
Insert the service cover at the end of the
line source compartment.
The small bolts mounted at the corners of the
cover must fit into the groove in the line
source compartment.