CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-51
Procedure Details
27.
Fasten the collision sensor to the end of
the line source compartment, mounting
one screw. See photo on the right.
28.
Assemble the spring and end stop as
shown in the photo.
29.
Attach the spring and end stop to the
collision detection shutter end cover.
30.
Mount the collision detection shutter end
cover onto the line source compartment.
Remember to mount the nut on the inside of
the line source compartment behind the spring
and end stop.
Proceed to remount covers, assemble the
transmission scanner and install it on the
detector:
31.
Install the cable connecting the line source
compartment to the bracket by reversing
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......