Setting Up FTP and Telnet
5-52
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
10.
Close the Add/Remove Programs dialog
by clicking the Close button in the upper
right-hand corner of the window.
11.
In the Control Panel, double-click
Administrative Tools.
12.
In the Administrative Tools window,
double-click Internet Services Manager.
13.
In the Internet Information Services
window, click the + next to the name of
the computer to unfold the branch of the
tree.
14.
Right-click Default FTP Site and select
Properties from the drop-down menu.
15.
In the Default FTP Site Properties dialog,
select the Home Directory tab at the top of
the window.
16.
On the Home Directory page, set Local
Path to
C:\
.
17.
Click OK.
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......