Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-27
Procedure Details
4.
To remove the large shipping bracket,
remove 2 bolts and 1 screw at each end of
the bracket.
8.3.3.3
Removing Gantry Shipping Pallet using Dolly Beams
Procedure Details
1.
Locate the 2 jack beams, the 2 wheel bars,
4 plastic protective feet, 4 screws M12
×
75 mm, and a ½” ratchet.
2.
Using 4 pcs. M12
×
75 mm socket head
cap screw and a 10 mm Allen wrench,
mount a jack beam in each end of the
gantry. The side with protective rubber
against the painted surface of the end bar.
3.
Apply a plastic protective foot under each
acme screw.
4.
Using a ½” ratchet, raise the gantry by
app. 1 inch (25 mm).
Remove bolts
Remove screw
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......