Software Update Procedures
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
7-19
Procedure Details
overwritten.
9.
When the software installation is com-
pleted, remove the software CD-ROM and
store it in a safe place along with the
backup disks.
10.
Start the CardioMD acquisition software
by double-clicking the CardioMD icon on
the acquisition PC’s desktop.
7.3.2
Performing a Complete Installation of the CardioMD Acquisition Software
7.3.2.1
Tools Required
•
CardioMD software installation CD, part number 9CSY0906.
Note
. A complete installation will overwrite all existing CardioMD acquisition software, detector
calibrations, non-default protocols etc. Make sure that up-to-date backup disks are available be-
fore performing a complete software installation. Instructions for performing backup and restore
of user data are provided in the Appendix of the CardioMD User’s Manual.
Note
. As a precaution, in case anything should go wrong, ensure that the previous version of the
software CD-ROM is available in order to be able to restore the original acquisition software ver-
sion.
7.3.2.2
Procedure
Before starting the acquisition PC software installation, check the release note on the CD-ROM.
You find the release note in the directory
Release notes
. The release note may provide infor-
mation that is more recent than this manual.
Logging On as Administrator
To perform a complete software installation, you need administrator privileges on the acquisition
PC. If your acquisition PC is running Windows XP, you already have administrator privileges. For
Windows 2000, proceed as follows:
Procedure Details
Close down the CardioMD application:
1.
If the CardioMD application is not dis-
playing the Persistence page, click the Exit
button to return to the Persistence page.
2.
On the Persistence page, click the Close
button in the upper right corner of the
CardioMD application window.
3.
Click the Windows Start button.
4.
Select
Shut Down
.
5.
In the drop-down list, select
Log off Car-
I, II
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......