Importing and Exporting Study Files
5-48
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
The Add a new node dialog appears.
7.
Select the FOLDER option and click
Select.
8.
Check off import/export capabilities from
the node as shown: No to Export and Yes
to Import.
Note
. If you want to both import and export
from the node check of Yes in both selections.
9.
To select the folder from where files will
be imported, click the … button.
10.
Ín the Drive drop-down, select the drive
where the folder containing the study files
is located.
The contents of the drive appear in the large
pane.
11.
Click on the folder’s name and click
Select.
The New Node dialog reappears.
12.
Click Save.
A new item appears in the Nodes
Configuration list.
13.
Click Exit.
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......