Diagnostics
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
5-27
Procedure Details
6.
On the acquisition PC’s desktop, right-
click the My Computer icon and select
Manage from the pop-up menu.
The Computer Management window appears.
7.
In the left-hand pane, select Device
Manager.
8.
In the right-hand pane, click the + next to
the entry Ports (Com & LPT).
9.
Check the COM port number stated next
to the item ATEN to USB Serial Bridge.
On a tower acquisition PC, you must use
COM1.
10.
Ensure that the correct COM port is
selected.
11.
Click OK to continue.
The COM Properties dialog appears.
12.
Set the following parameters for the serial
interface:
Baud rate: 19200
Data bits: 8
Parity: None
Stop bits: 1
Flow control: None.
13.
Select OK to continue.
The HyperTerminal main window appears.
III
III
I, II
III
I, II
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......