Diagnostics
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
5-53
Procedure Details
18.
Close the Internet Information Services
window.
19.
Leave the Administrative Tools window
and the Control panel open.
5.8.1.3
Setting Up Telnet
In this procedure, you set up a Telnet server on the acquisition PC. When this is accomplished, a
remote user can log into the acquisition PC via Telnet.
Procedure Details
1.
In the Administrative Tools window,
double-click Telnet Server Administration.
A command window appears.
2.
Enter
3
[Enter] to Display/change registry
settings.
3.
Enter
7
[Enter] to select NTLM.
4.
When prompted, enter
y
[Enter] to
confirm.