NEMA Multiple Window Spatial Registration (MWSR)
10-14
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
After the last 400 kilocounts the result pops
up.
16.
Note the result.
The MWSR specification is
≤
2.0 mm.
17.
Remove the GEOCAL mask and the re-
maining lead pig from the detector and
mount the collimator.
18.
Click Exit to return to the NEMA menu.
Note
. To measure the other detector, place the CardioMD system in the patient load position be-
fore repeating the procedure outlined above.