CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-73
Procedure Details
4.
Carefully push the slide bearings into
position underneath the line source
housing.
5.
Use a 2.5 mm Allen key to mount the 8
M3
×
30 Allen screws, securing the line
source housing to the slide bearings.
Note
. Before fastening the screws, check that
the line source housing is perpendicular to the
rails, e.g. by aligning one of its sides with
screw heads in the two rails.
If required, turn the motor shaft by hand to
translate the line souce housing to a more
convenient position inside the compartment.
The photo on the right shows the line source
housing installed on the slide bearings and
with all screws mounted.
6.
Finally, remove the two transportation
screws securing the line source housing
shutter mechanism.
Transportation screws
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......