Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-51
Procedure Details
4.
Locate 4 large steel washers in the
installation kit.
5.
Place two washers on the pivot shaft.
6.
Assure that the cable is routed through the
arm and escapes through the bottom of the
arm.
7.
Route the cable through the split ring of
the wall mounting plate, assuring that the
cable grommet is below the split ring.
8.
Place the arm between the two rings on the
wall mounting plate.
9.
Place the remaining two large steel
washers on top of the arm, under the top
ring of the wall mounting plate.