Installation
8-36
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
17 mm wrench.
8.3.3.14
Mounting the Table Console to the Gantry Frame
Procedure Details
1.
Using the long bolts as guides, push the
table base in position. A key in the gantry
frame will guide the base the last ¼”
(5 mm).
2.
Remove the long screws and replace them
with M10
×
20 mounting screws, one at
the time.
3.
Using an 8 mm Allen wrench, tighten the
screws and loosen each screw again by
approximately half a turn of the wrench.
4.
Lower the rear end of the table base and
remove the jack beam.
8.3.3.15
Mounting Anchors for Table Console
Procedure Details
1.
Using a 10 mm Allen key, mount 2 M12
×
30 anchor screws with washers at the rear
end of the table. Tighten the bolts.
2.
Also tighten the 2 anchor bolts at the front
end of the gantry.
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......