Installation
8-74
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
1.
Cut the cable protectors to the required
lengths.
2.
Open the membrane on the base of the
protector to slide in the cables.
If an insulator is required to separate the
power line from the network cable:
3.
Slice one of the columns inside the rubber
cable protector to place one of the two
cables in the side compartment.
WARNING
Trip hazard: Failing to protect and secure the gantry-PC interface cable may
cause cables to be damaged and persons to trip over the cable.
Slice or cut this column
within the rubber cable
protector to isolate the
power line from the network
cable.