Spare Parts
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
9-37
Procedure Details
4.
Attach the collimator number label (I or
II) onto the collimator in the location
indicated in the diagram on the right.
5.
After replacing a collimator, you must
acquire:
-
A new center of rotation correction
table
-
New extrinsic uniformity calibration
tables.
Refer to the Operator’s Manual Chapter 5
Maintenance
for procedures for performing
center of rotation calibration and extrinsic
uniformity calibration.
I or II label position
Slanted part
Handle
Side towards
patient
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......