Calibration
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
4-35
Procedure Details
5.
Use the hand controller to move the detec-
tors in the X direction until the distance
between the front edge of the moveable
gantry horizontal base and the back edge
of the front gantry frame is exactly
234 mm (9.2”).
To reach the Y calibration position:
6.
Move the detectors vertically until the
distance between the top surface of the de-
tector rotate arm and the top edge of the
vertical direction opening is exactly
790 mm (31 3/32”).
See the photo on the right.
Now, proceed to perform the calibration:
7.
On the safety board, press ZERO_ENA
and ZERO_R simultaneously to calibrate
the rotate absolute encoder.
8.
On the safety board, press ZERO_ENA
and ZERO_V simultaneously to calibrate
the vertical absolute encoder.
9.
On the safety board, press ZERO_ENA
and ZERO_H simultaneously to calibrate
the horizontal absolute encoder.
10.
Reset both motion controllers by setting
switch 10 in ON position (upwards) on
ZERO_ENA
ZERO_R
ZERO_H
ZERO_V
Содержание CARDIOMD
Страница 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 207: ...Installation 1MAN0158 H11 04 2005 CSIP Level 1 CardioMD Service Manual 8 5 Figure 8 2 System footprint with short table ...
Страница 232: ...Installation 8 30 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 Figure 8 17 Installation template ...
Страница 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Страница 465: ......